19510272-V2.doc 11.07.13 English Operating manual Refrigerated / Heating Circulator F12-ED F25-ED F26-ED JULABO Gm
Operator responsibility – Safety recommendations 10 Safety installations according to IEC 61010-2-010: Excess temperature protection adjustable fro
ED 11 Operating instructions 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed, warnings are
Safety notes for the user 12 Do not stay in the area below the unit. Make sure you read and understand all instructions and safety precautions l
ED 13 Caution: The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice. We do not know which substances are c
Operating controls and functional elements 14 4. Operating controls and functional elements Front view Rear view 1a 1b Mains power swit
ED 15 5 Control indicator –Heating 6 Control indicator – Alarm 7 Adjustable excess temperature protection according to IEC 61010-2-010 8a 15AM
Preparations 16 5. Preparations 5.1. Installation Place the unit on an even surface on a pad made of non-flammable material. Set select
ED 17 5.2. Bath fluids Caution: No liability for use of other bath fluids! Do not use flammable bath fluids! Water: The quality of water depends
Preparations 18 5.3. Filling / Draining Notice: Filling Check the filling level of the bath fluid from time to time. Pump and heater must alway
ED 19 5.4. Temperature application to external systems Caution: Securely attach all tubing to prevent slipping. If the circulator is operated witho
2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been design
Preparations 20 Warning: Tubing: At high working temperatures the tubing used for temperature application and cooling water supply represents a da
ED 21 6. Operating procedures 6.1. Power connection Caution: Only connect the unit to a power socket with earthing contact (PE – protective earth
Operating procedures 22 Caution: If the circulator is turned off with the mains switch (1a), the refrigerating unit is not switched off simultaneo
ED 23 6.4. Setting the temperatures Factory setting: 25 °C Setting can be carried out in the start/stop condition. 1. Press one of the keys f
Troubleshooting guide / Error messages 24 7. Troubleshooting guide / Error messages + Whenever the microprocessor electronics registers a failure,
ED 25 15AMP Circulator: Mains circuit breakers (resettable) 15 A Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm The mains fuses on the rear of th
Cleaning / repairing the unit 26 To maintain the full cooling performance, clean the condenser from time to time. Switch off the unit, disconnect
ED 3 TABLE OF CONTENTS Operating manual ...
Operating manual 4 Operating manual 1. Intended use JULABO circulators have been designed to control the temperature of specific fluids in a bath tan
ED 5 Persons: The operator is responsible for the qualification of the personnel operating the units. The operator should be constantly informed ab
Operator responsibility – Safety recommendations 6 When using hazardous materials or materials that could become hazardous, the operator must affix th
ED 7 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following European guidelines: Dire
Operator responsibility – Safety recommendations 8 2.4. Technical specifications F12-ED F25-ED Working temperature range °C -20 ... 100 -28 .
ED 9 F26-ED Working temperature range °C -28 ... 100 Temperature stability °C ±0,03 Temperature selection digital Temperature in